Developing Metrological Assurance for Measurements of Optical Density of X-Ray Images in a Range of 0.2 to 6.3


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

We present results of developing metrological assurance for the measurements of the optical density of an X-ray image based on a set of optical density measures in the range of 0.2 to 6.3. The set has been designed to render one unit of diffuse optical density for densitometers used to measure optical density and estimate the size of defects on X-ray images when inspecting nuclear power facilities, oil and gas enterprises, railway transport, metallurgical and mechanical-engineering plants, etc. Results of calibrating the set of optical density measures with the GET 206-2016 State Primary Standard of the unit of optical density, developed and in operation at VNIIOFI, are presented.

About the authors

A. V. Koldashov

All-Russian Research Institute for Optical and Physical Measurements (VNIIOFI)

Email: marchenko@vniiofi.ru
Russian Federation, Moscow, 119361

S. N. Marchenko

All-Russian Research Institute for Optical and Physical Measurements (VNIIOFI)

Author for correspondence.
Email: marchenko@vniiofi.ru
Russian Federation, Moscow, 119361

V. I. Kapustin

OOO AVKONT

Email: marchenko@vniiofi.ru
Russian Federation, Moscow, 119334


Copyright (c) 2019 Pleiades Publishing, Ltd.