A Device for Thermal Treatment of Luminescence Ionizing-Radiation Detectors Intended for Radiation Nondestructive Testing


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

The structure and principle of operation of a heating unit intended for the thermal and radiation-thermal treatment of luminescence ionizing-radiation detectors aimed at modifying their dosimetric characteristics are described. The unit provides for linear heating of detectors in the range of 50–700°C at a given rate from 1 to 15°C/s, as well as for isothermal exposure of detectors. A description of unit operation is given as well as the results of trial tests that prove a high accuracy in reproducing the temperature and heating rate, acceptable for problems being solved.