The Design of an Open MRI 4-Channel Receive-Only Phased Array Knee Coil


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

In this work, a new prototype 4-channel receive-only phased array (PA) knee coil for open magnetic resonance imaging (MRI) system was designed and constructed. The knee coil was used in a series of pilot MRI studies on a homogenous cylindrical phantom and human knee. From the simulation and experimental results, it is shown that both the arrangement of the coil elements and the width of the coil elements within this new coil array are critical design factors for the coil array to work optimally in an open MRI system. Experimental results showed that a uniform phantom image can be acquired by the prototype phased array knee coil and human knee images further demonstrated the feasibility of the proposed method. The proposed method is appropriate for the design of other open MRI phased array coils.

作者简介

Xiaowei He

School of Information Sciences and Technology, Northwest University

编辑信件的主要联系方式.
Email: hexw@nwu.edu.cn
中国, Xi’an

Renbo Yuan

School of Information Sciences and Technology, Northwest University

Email: hexw@nwu.edu.cn
中国, Xi’an

Bing Li

Jiangsu Magspin Instrument Co. Ltd; School of Information Technology and Electrical Engineering, University of Queensland

Email: hexw@nwu.edu.cn
中国, Kunshan; Brisbane

Yuqing Hou

School of Information Sciences and Technology, Northwest University

Email: hexw@nwu.edu.cn
中国, Xi’an

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer-Verlag Wien, 2016