The Design of an Open MRI 4-Channel Receive-Only Phased Array Knee Coil


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

In this work, a new prototype 4-channel receive-only phased array (PA) knee coil for open magnetic resonance imaging (MRI) system was designed and constructed. The knee coil was used in a series of pilot MRI studies on a homogenous cylindrical phantom and human knee. From the simulation and experimental results, it is shown that both the arrangement of the coil elements and the width of the coil elements within this new coil array are critical design factors for the coil array to work optimally in an open MRI system. Experimental results showed that a uniform phantom image can be acquired by the prototype phased array knee coil and human knee images further demonstrated the feasibility of the proposed method. The proposed method is appropriate for the design of other open MRI phased array coils.

Sobre autores

Xiaowei He

School of Information Sciences and Technology, Northwest University

Autor responsável pela correspondência
Email: hexw@nwu.edu.cn
República Popular da China, Xi’an

Renbo Yuan

School of Information Sciences and Technology, Northwest University

Email: hexw@nwu.edu.cn
República Popular da China, Xi’an

Bing Li

Jiangsu Magspin Instrument Co. Ltd; School of Information Technology and Electrical Engineering, University of Queensland

Email: hexw@nwu.edu.cn
República Popular da China, Kunshan; Brisbane

Yuqing Hou

School of Information Sciences and Technology, Northwest University

Email: hexw@nwu.edu.cn
República Popular da China, Xi’an

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer-Verlag Wien, 2016