An X-Band Crossed-Loop EPR Resonator


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

A copper X-band (9.22 GHz) cross-loop resonator has been constructed for use with 4 mm sample tubes. The Q for the two resonators is 380 and 350, respectively. The resonator efficiency is about 1 G per square root of watt. Operation has been demonstrated with measurement of T1 by saturation recovery for samples of coal and an immobilized nitroxide.

作者简介

George Rinard

School of Engineering and Computer Science and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
美国, Denver, CO, 80208

Richard Quine

School of Engineering and Computer Science and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
美国, Denver, CO, 80208

Joseph McPeak

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
美国, 2101 E. Wesley Ave., Denver, CO, 80208

Laura Buchanan

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
美国, 2101 E. Wesley Ave., Denver, CO, 80208

Sandra Eaton

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
美国, 2101 E. Wesley Ave., Denver, CO, 80208

Gareth Eaton

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

编辑信件的主要联系方式.
Email: geaton@du.edu
ORCID iD: 0000-0001-7429-8469
美国, 2101 E. Wesley Ave., Denver, CO, 80208

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer-Verlag GmbH Austria, 2017