At the Test Laboratory of CITO five modern devices for transosseous osteosynthesis were studied to evaluate the fixation stability in comminuted fractures. Firmness was defined in compression, twisting as well as frontal and sagittal bends. In all planes the highest level of fixation stability was found in RAPFIS device. Design peculiarity of that device provided the additional increase of fixation stability by roads insertion into bone fragments at an angle and diminution of the «bonefixative» distance (console) that almost 3 times improve the fixation stability.

About the authors

A. I. Gorodnichenko

Medical Center of the Administrative Department of the President of the Russian Federation;, Central Institute of Traumatology and Orthopedics. N.N. Priorov; Research and Production Association "UVICOM"

Author for correspondence.
Email: info@eco-vector.com
Russian Federation, Moscow; Moscow; Mytishchi

N. S. Gavryushenko

Medical Center of the Administrative Department of the President of the Russian Federation;, Central Institute of Traumatology and Orthopedics. N.N. Priorov; Research and Production Association "UVICOM"

Email: info@eco-vector.com
Russian Federation, Moscow; Moscow; Mytishchi

M. E. Kazakov

Medical Center of the Administrative Department of the President of the Russian Federation;, Central Institute of Traumatology and Orthopedics. N.N. Priorov; Research and Production Association "UVICOM"

Email: info@eco-vector.com
Russian Federation, Moscow; Moscow; Mytishchi

V. M. Kernichansky

Medical Center of the Administrative Department of the President of the Russian Federation;, Central Institute of Traumatology and Orthopedics. N.N. Priorov; Research and Production Association "UVICOM"

Email: info@eco-vector.com
Russian Federation, Moscow; Moscow; Mytishchi

References

  1. Alonso J. //Clin. Orthop. — 1989. — N 241. — Р. 83-88.
  2. Broekhuizen Т.Н., Вохта Н. et al. //Injury. — 1990.— Vol. 21. — P. 145-151.
  3. Chao EYS. //J. Biomech. — 1982. — N 12. — P. 971- 983.
  4. Chao EYS. //Orthopedics. — 1988. — N 11. — P. 1057-1069.
  5. Chevalley F., Amsutz Ch., Bally A. //J. Orthop. Surg.— 1993. — N 1. — P. 1-7.
  6. Finlay J.B., Moroz T.K. //J. Bone Jt Surg. — 1987. — Vol. 69A. — P. 734-744.
  7. McCoy M.T., Chao EYS. //Clin. Orthop. — 1983. — N 180. — P. 23-33.

Supplementary files

Supplementary Files
Action
1. JATS XML
2. Fig. 1. Conditions for testing rod devices on a bone simulator. 1 - bone simulator; 2 - rod; 3 — apparatus for external fixation.

Download (17KB)
3. Fig. 2. Rigidity of various configurations of the RAPFIS apparatus under compression. I - console 60 mm, II - console 35 mm, III - console 35 mm, rods at an angle.

Download (16KB)

Copyright (c) 1999 Eco-Vector



Согласие на обработку персональных данных

 

Используя сайт https://journals.rcsi.science, я (далее – «Пользователь» или «Субъект персональных данных») даю согласие на обработку персональных данных на этом сайте (текст Согласия) и на обработку персональных данных с помощью сервиса «Яндекс.Метрика» (текст Согласия).