Intrinsic noise of seismometers at frequencies from 0.01 to 0.1 Hz
- Авторы: Shevchenko Y.V.1, Yakovenko V.V.1
-
Учреждения:
- Metrological Service Group, Kamchatka Branch, Geophysical Service
- Выпуск: Том 53, № 1 (2017)
- Страницы: 70-80
- Раздел: Article
- URL: https://journals.rcsi.science/0747-9239/article/view/177380
- DOI: https://doi.org/10.3103/S074792391701008X
- ID: 177380
Цитировать
Аннотация
The paper presents a comparative analysis of the intrinsic noises of seismometers operating at Kamchatka seismic stations: CMG-3TB (GURALP), CMG-5TD (GURALP), CMG-6TD (GURALP), GMS-AC-73i HHV (GeoSIG), SM3os (GEOTECH+), S5Sos (KF GS), and SM3kv. It is shown that the dynamic range of the seismometers is narrower at low frequencies in the area of weak signals.