Hidden Uncertainties of Temperature Measurements in the Calibration of a Pyrometer


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

We consider some (earlier neglected) uncertainties of temperature measurements that may appear in the course of calibration of radiation thermometers (pyrometers) aimed at measuring temperatures within the range 1234.93–3273.15 K. It is shown that one of these uncertainties is connected with the difference between the spectral transfer functions of the interference filter of the calibrated pyrometer and the monochromator of a working temperature standard. The other uncertainty is caused by the asymmetry of the spectral transfer function of the pyrometer about the central wavelength of its passband.

作者简介

V. Khodunkov

Mendeleev All-Russia Research Institute of Metrology (VNIIM)

编辑信件的主要联系方式.
Email: walkerearth@mail.ru
俄罗斯联邦, St. Petersburg