Determining the Quantum Characteristics of a Photon Counter Using Power-Calibrated Laser Radiation
- Авторы: Kovalev A.A.1, Liberman A.A.1, Mikryukov A.S.1, Moskalyuk S.A.1, Ulanovskii M.V.1
-
Учреждения:
- All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
- Выпуск: Том 58, № 11 (2016)
- Страницы: 1244-1249
- Раздел: Article
- URL: https://journals.rcsi.science/0543-1972/article/view/245573
- DOI: https://doi.org/10.1007/s11018-016-0878-4
- ID: 245573
Цитировать
Аннотация
An experimental setup was devised allowing high-accuracy radiation power measurements in the range of the order of 10−18–10−15 W. The use of a calibrated Fresnel prismatic attenuator in the design minimizes the error in determining the mean number of photos in each radiation pulse. Results are presented for measurements of the positive measure operator for a Hamamatsu H7422-40 photoelectric multiplier.
Об авторах
A. Kovalev
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Автор, ответственный за переписку.
Email: mrsergik@mail.ru
Россия, Moscow
A. Liberman
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Россия, Moscow
A. Mikryukov
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Россия, Moscow
S. Moskalyuk
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Россия, Moscow
M. Ulanovskii
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Россия, Moscow
Дополнительные файлы