Measurement of the Spectral Irradiance of Deuterium Lamps in the Range of Wavelengths 200–400 nm Using a High-Temperature Black Body Model


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

A procedure is presented for measurements of the spectral irradiance of deuterium lamps in the wavelength range 200–400 nm with unit reproduction using a high-temperature blackbody model. It is shown that the expanded measurement uncertainty varies between 1.6–6.9%. The measurement results were confirmed with calibrations between VNIIOFI (Russia) and PTB (Germany). The divergence of the results obtained by the two laboratories does not exceed 2%.

Sobre autores

B. Khlevnoy

All-Russia Research Institute of Optophysical Measurements (VNIIOFI)

Autor responsável pela correspondência
Email: khlevnoy-m4@vniiofi.ru
Rússia, Moscow

M. Solodilov

All-Russia Research Institute of Optophysical Measurements (VNIIOFI)

Email: khlevnoy-m4@vniiofi.ru
Rússia, Moscow

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer Science+Business Media, LLC, part of Springer Nature, 2019