Measurement of Magnetic Susceptibility and Curie Constants of Colloidal Solutions in Ferrofluid Cells by the Nuclear Magnetic Resonance Method
- Авторлар: Davydov V.V.1,2,3, Myazin N.S.1
-
Мекемелер:
- St. Petersburg Polytechnic University
- St. Petersburg State Telecommunications University
- All-Russia Research Institute of Phytopathology
- Шығарылым: Том 60, № 5 (2017)
- Беттер: 491-496
- Бөлім: Electromagnetic Measurements
- URL: https://journals.rcsi.science/0543-1972/article/view/246171
- DOI: https://doi.org/10.1007/s11018-017-1223-2
- ID: 246171
Дәйексөз келтіру
Аннотация
A new method is proposed for determining the magnetic susceptibility and Curie constants of various colloidal solutions in sealed glass volumes (ferrofluid cells) with an error of no more than 1.0% even with a small thickness of the layer of solution. It is shown that intensive stirring of the solution is necessary for excluding changes in the Curie constants as the cell temperature increases.