Calibration and Validation of Dynamic Force Transducers
- 作者: Prilepko M.Y.1
-
隶属关系:
- All-Russia Research Institute of Metrological Service (VNIIMS)
- 期: 卷 60, 编号 7 (2017)
- 页面: 691-695
- 栏目: Nanometrology
- URL: https://journals.rcsi.science/0543-1972/article/view/246220
- DOI: https://doi.org/10.1007/s11018-017-1255-7
- ID: 246220
如何引用文章
详细
The urgent need for a procedure to use in validation of dynamic force transducers is considered. A model of the dynamic interaction of the elements of a dynamic force transducer in the course of calibration by the test mass method is presented. Mathematical expressions that describe the calibration process are presented. A schematic diagram of a calibration plant that would be most appropriate for dynamic force transducers is proposed.