Applicability of an Alternative Method for Sterility Testing of Medicinal Preparations
- Авторы: Roshchina M.1, Gunar O.1, Sakhno N.1
-
Учреждения:
- Scientific Centre for Expert Evaluation of Medicinal Products, Ministry of Health of the Russian Federation
- Выпуск: Том 51, № 11 (2018)
- Страницы: 1045-1048
- Раздел: Article
- URL: https://journals.rcsi.science/0091-150X/article/view/244906
- DOI: https://doi.org/10.1007/s11094-018-1737-1
- ID: 244906
Цитировать
Аннотация
The aim of the present work was to examine the applicability of an alternative method for detecting CO2 using an automated BacT/ALERT 3D colorimetric microbiological detection system and analytical parameters for sterility testing of medicinal preparations (MPs) that were selected, justified, and harmonized with leading global pharmacopoeias. The tests used microorganisms and samples of 38 MPs and compared the results with defined criteria for pharmaceutical microbiology. The experimental results showed that the alternative method did not cause discrepancies and was suitable for MPsterility testing. However, the lack of regulations for validation standard procedures in the RF currently limits replacing routine methods by alternative ones.
Об авторах
M. Roshchina
Scientific Centre for Expert Evaluation of Medicinal Products, Ministry of Health of the Russian Federation
Email: chem@folium.ru
Россия, 8/2 Petrovskii Blvd., Moscow, 127051
O. Gunar
Scientific Centre for Expert Evaluation of Medicinal Products, Ministry of Health of the Russian Federation
Email: chem@folium.ru
Россия, 8/2 Petrovskii Blvd., Moscow, 127051
N. Sakhno
Scientific Centre for Expert Evaluation of Medicinal Products, Ministry of Health of the Russian Federation
Email: chem@folium.ru
Россия, 8/2 Petrovskii Blvd., Moscow, 127051