Assessment of the Effects of Deformation of the Fibrous Ring of the Aortic Valve on the Hydrodynamic Characteristics of Transcatheter Prostheses In Vitro
- Autores: Boykov A.A.1, Tarkova A.R.1, Kretov E.I.1, Prokhorikhin A.A.1, Malaev D.U.1
-
Afiliações:
- Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
- Edição: Volume 53, Nº 4 (2019)
- Páginas: 275-279
- Seção: Article
- URL: https://journals.rcsi.science/0006-3398/article/view/236206
- DOI: https://doi.org/10.1007/s10527-019-09925-w
- ID: 236206
Citar
Resumo
Questions of the occurrence of transprosthesis regurgitation after transcathether prosthetization of the aortic valve are discussed. We present results from hydrodynamic tests of prototype transcatheter aortic valve prostheses with different methods of attachment of the valve leaflets in conditions of deformation of the fibrous ring. Prototype prostheses with attachment of the leaflets of the valve apparatus above the seating zone of the frame of the prosthesis demonstrated optimal hydrodynamic characteristics.
Sobre autores
A. Boykov
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Autor responsável pela correspondência
Email: fimarik@gmail.com
Rússia, Novosibirsk
A. Tarkova
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Email: fimarik@gmail.com
Rússia, Novosibirsk
E. Kretov
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Email: fimarik@gmail.com
Rússia, Novosibirsk
A. Prokhorikhin
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Email: fimarik@gmail.com
Rússia, Novosibirsk
D. Malaev
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Email: fimarik@gmail.com
Rússia, Novosibirsk
Arquivos suplementares