Measurement of Characteristics of Cryoprobes under Different Operation Modes
- Autores: Zherdev A.A.1, Shakurov A.V.1, Pushkarev A.V.1, Burkov I.A.1, Antonov E.A.1, Vasilev A.O.2, Belozerov A.G.1
-
Afiliações:
- Bauman Moscow State Technical University
- Moscow State University of Medicine and Dentistry
- Edição: Volume 50, Nº 5 (2017)
- Páginas: 344-347
- Seção: Article
- URL: https://journals.rcsi.science/0006-3398/article/view/234744
- DOI: https://doi.org/10.1007/s10527-017-9652-4
- ID: 234744
Citar
Resumo
Experimental time curves of temperature distribution over the length of the working surface of a minimally invasive cryoprobe were measured for various modes of operation (cooling or heating). The cooldown time was determined. A custom-built experimental installation allowed different types of minimally invasive cryoprobes to be tested. The results of the experiments can be used in planning cryosurgery using commercially available cryoprobes, as well as for testing the efficiency of newly developed cryosurgical instruments.
Sobre autores
A. Zherdev
Bauman Moscow State Technical University
Autor responsável pela correspondência
Email: azherdev@bmstu.ru
Rússia, Moscow
A. Shakurov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Rússia, Moscow
A. Pushkarev
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Rússia, Moscow
I. Burkov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Rússia, Moscow
E. Antonov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Rússia, Moscow
A. Vasilev
Moscow State University of Medicine and Dentistry
Email: azherdev@bmstu.ru
Rússia, Moscow
A. Belozerov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Rússia, Moscow
Arquivos suplementares