Measurement of Characteristics of Cryoprobes under Different Operation Modes
- Авторы: Zherdev A.A.1, Shakurov A.V.1, Pushkarev A.V.1, Burkov I.A.1, Antonov E.A.1, Vasilev A.O.2, Belozerov A.G.1
-
Учреждения:
- Bauman Moscow State Technical University
- Moscow State University of Medicine and Dentistry
- Выпуск: Том 50, № 5 (2017)
- Страницы: 344-347
- Раздел: Article
- URL: https://journals.rcsi.science/0006-3398/article/view/234744
- DOI: https://doi.org/10.1007/s10527-017-9652-4
- ID: 234744
Цитировать
Аннотация
Experimental time curves of temperature distribution over the length of the working surface of a minimally invasive cryoprobe were measured for various modes of operation (cooling or heating). The cooldown time was determined. A custom-built experimental installation allowed different types of minimally invasive cryoprobes to be tested. The results of the experiments can be used in planning cryosurgery using commercially available cryoprobes, as well as for testing the efficiency of newly developed cryosurgical instruments.
Об авторах
A. Zherdev
Bauman Moscow State Technical University
Автор, ответственный за переписку.
Email: azherdev@bmstu.ru
Россия, Moscow
A. Shakurov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Россия, Moscow
A. Pushkarev
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Россия, Moscow
I. Burkov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Россия, Moscow
E. Antonov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Россия, Moscow
A. Vasilev
Moscow State University of Medicine and Dentistry
Email: azherdev@bmstu.ru
Россия, Moscow
A. Belozerov
Bauman Moscow State Technical University
Email: azherdev@bmstu.ru
Россия, Moscow
Дополнительные файлы