Mathematical Modeling of Hemolysis in Pulsatile Blood Pumps


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

The paper presents the results of calculation of the normalized index of hemolysis using three different approaches based on mathematical modeling of a pulsatile circulatory support system with a stroke volume of 30 cm3. The suggested approaches are shown to be useful for estimating the hemolysis level in pulsatile circulatory support systems at various stages of the design process.