Simulation and Control of the H−Q Curves of the Sputnik Portable Ventricular Assist Device
- Authors: Petukhov D.S.1
-
Affiliations:
- National Research University of Electronic Technology
- Issue: Vol 50, No 6 (2017)
- Pages: 429-432
- Section: Article
- URL: https://journals.rcsi.science/0006-3398/article/view/234834
- DOI: https://doi.org/10.1007/s10527-017-9671-1
- ID: 234834
Cite item
Abstract
A previously suggested approach to identification of pumping states of a rotary blood pump was verified using the head−flow curves of the Sputnik ventricular assist device. The curves were obtained using a mock circulatory loop under static conditions. The obtained results can be used for implementing a rotary blood pump control strategy intended to maintain the required flow rate and to identify the associated pumping states.
About the authors
D. S. Petukhov
National Research University of Electronic Technology
Author for correspondence.
Email: petuhov@bms.zone
Russian Federation, Zelenograd, Moscow
Supplementary files
