Flow Sensor in Devices for Autotransfusion


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

This article describes an original multi-channel flow sensor in the TRANSFEL autotransfusion device. The flow sensor in the TRANSFEL device is based on digital processing of multiple CRGB channels. The article presents the position of the sensor in the TRANSFEL device and its principle of operation and structure. Experimental results demonstrate the advantage of this sensor over its analogues since it more precisely defines the composition of the fluid going into the waste line.

About the authors

I. N. Selyanko

CJSC NTTs ELINS, Zelenograd

Author for correspondence.
Email: selyanko@elins.ru
Russian Federation, Moscow

E. V. Medvedeva

CJSC NTTs ELINS, Zelenograd

Email: selyanko@elins.ru
Russian Federation, Moscow

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2016 Springer Science+Business Media New York