Effect of feedback loop on the resolution of acousto-optic spectrometer
- Авторы: Mantsevich S.N.1, Balakshy V.I.1, Kuznetsov Y.I.1
-
Учреждения:
- Faculty of Physics
- Выпуск: Том 24, № 2 (2016)
- Страницы: 135-141
- Раздел: Acousto-Optic Effect
- URL: https://journals.rcsi.science/1541-308X/article/view/217536
- DOI: https://doi.org/10.3103/S1541308X16020084
- ID: 217536
Цитировать
Аннотация
An acousto-optic spectrometer based on a collinear acousto-optic cell with feedback loop has been theoretically and experimentally investigated. It is shown that the introduction of optoelectronic feedback affects to a great extent the characteristics of collinear acousto-optic interaction and makes it possible to increase significantly the accuracy of determining the wavelength of incident optical signal.
Об авторах
S. Mantsevich
Faculty of Physics
Автор, ответственный за переписку.
Email: snmantsevich@yahoo.com
Россия, Leninskie Gory 1, str. 2, Moscow, 119991
V. Balakshy
Faculty of Physics
Email: snmantsevich@yahoo.com
Россия, Leninskie Gory 1, str. 2, Moscow, 119991
Yu. Kuznetsov
Faculty of Physics
Email: snmantsevich@yahoo.com
Россия, Leninskie Gory 1, str. 2, Moscow, 119991
Дополнительные файлы