Effects of scanning speed on the laser beam profile measurements by vibrating wire
- 作者: Arutunian S.G.1, Harutynyan G.S.1, Choe D.2, Chung M.2, Lazareva E.G.1, Margaryan A.V.1
-
隶属关系:
- Alikhanyan National Scientific Laboratory
- Ulsan National Institute of Science and Technology
- 期: 卷 52, 编号 4 (2017)
- 页面: 366-374
- 栏目: Article
- URL: https://journals.rcsi.science/1068-3372/article/view/228517
- DOI: https://doi.org/10.3103/S1068337217040090
- ID: 228517
如何引用文章
详细
For measuring laser beam profiles, a vibrating wire monitor (VWM) has been introduced. The measurements were carried out at different speeds of scan. Preliminary estimates were made for the calculation of the VWM response times with respect to the thermal losses along the wire, and radiative and convective losses. These estimates, however, do not determine the difference between the beam profile and the frequency response of the VWM for a given scan rate. To evaluate the reliability of the frequency response of the VWM, comparisons between forward and reverse beam scans at different speeds have been used. The results of these scans are used to correct the thermal inertia in the frequency response of the VWM.
作者简介
S. Arutunian
Alikhanyan National Scientific Laboratory
Email: ella.lazareva@yerphi.am
亚美尼亚, Yerevan
G. Harutynyan
Alikhanyan National Scientific Laboratory
Email: ella.lazareva@yerphi.am
亚美尼亚, Yerevan
D. Choe
Ulsan National Institute of Science and Technology
Email: ella.lazareva@yerphi.am
韩国, Ulsan
M. Chung
Ulsan National Institute of Science and Technology
Email: ella.lazareva@yerphi.am
韩国, Ulsan
E. Lazareva
Alikhanyan National Scientific Laboratory
编辑信件的主要联系方式.
Email: ella.lazareva@yerphi.am
亚美尼亚, Yerevan
A. Margaryan
Alikhanyan National Scientific Laboratory
Email: ella.lazareva@yerphi.am
亚美尼亚, Yerevan
补充文件