Changes in the Morphological and Mechanical Characteristics of Bone Implants upon Radiation Sterilization


Цитировать

Полный текст

Открытый доступ Открытый доступ
Доступ закрыт Доступ предоставлен
Доступ закрыт Только для подписчиков

Аннотация

Radiation sterilization of bone implants can lead to substantial morphological changes in them and the degradation of their physical and mechanical properties, depending on the magnitude of the absorption dose. The structure of bone samples changes starting at doses of 15 kGy. Raising the dose to 25–50 kGy affects the mechanical characteristics and osteinductive properties of bone implants. Data on the morphological and mechanical changes, and their analysis and systematization, are the scientific basis for developing modern requirements and practical recommendations for further improvement of the radiation technologies for the sterilization of biological tissues.

Об авторах

V. Rozanov

Moscow State University; National Research Institute for the Development of Medicinal and Aromatic Plants

Автор, ответственный за переписку.
Email: vrozanov@mail.ru
Россия, Moscow, 119991; Moscow, 117216

I. Matveychuk

National Research Institute for the Development of Medicinal and Aromatic Plants

Email: vrozanov@mail.ru
Россия, Moscow, 117216

A. Chernyaev

Moscow State University

Email: vrozanov@mail.ru
Россия, Moscow, 119991

N. Nikolaeva

Northeastern Federal University

Email: vrozanov@mail.ru
Россия, Yakutsk, Republic of Sakha, 677000

Дополнительные файлы

Доп. файлы
Действие
1. JATS XML

© Allerton Press, Inc., 2019

Согласие на обработку персональных данных

 

Используя сайт https://journals.rcsi.science, я (далее – «Пользователь» или «Субъект персональных данных») даю согласие на обработку персональных данных на этом сайте (текст Согласия) и на обработку персональных данных с помощью сервиса «Яндекс.Метрика» (текст Согласия).