New possibilities and some artifacts of the cathodoluminescent mode in scanning electron microscopy

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

Lightning inside the chamber of a scanning electron microscope (SEM), caused by electrons being scattering from a sample (and parts of the chamber), is observed and analyzed. These electrons generate the luminescence in a Thornly–Everhart collector. This parasitic effect (artifact) must be considered and eliminated in all experiments with the cathodoluminescent (CL) mode of SEM. A new technique for measuring surface potential on dielectric samples is proposed. It is based on variations in the CL signal during electron irradiation of a sample in SEM.

Авторлар туралы

S. Zaitsev

Faculty of Physics

Email: rau@phys.msu.ru
Ресей, Moscow, 119991

S. Kupreenko

Faculty of Physics

Email: rau@phys.msu.ru
Ресей, Moscow, 119991

A. Luk’yanov

Faculty of Physics

Email: rau@phys.msu.ru
Ресей, Moscow, 119991

E. Rau

Faculty of Physics

Хат алмасуға жауапты Автор.
Email: rau@phys.msu.ru
Ресей, Moscow, 119991

A. Tatarintsev

Faculty of Physics

Email: rau@phys.msu.ru
Ресей, Moscow, 119991

A. Khaidarov

Faculty of Physics

Email: rau@phys.msu.ru
Ресей, Moscow, 119991

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Allerton Press, Inc., 2016