Eliminating saturation-type distortions in ultrasonic signals by the least-squares and Papoulis–Gerchberg algorithms


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

Situations may occur in ultrasonic testing in which the values of recorded echo-signals exceed the dynamic range of the receiving amplifier and defectoscope’s analog-to-digital converter. As a result, high-amplitude signals are subjected to a cutoff (clipping) operation and drop in amplitude, thus introducing error in the estimation of reflector dimensions. A declipping method that is based on the Papoulis–Gerchberg algorithm has been suggested and compared against a declipping technique that uses least squares. In numerical and model experiments, the Papoulis–Gerchberg algorithm has demonstrated more stable operation than the least squares for noisy signals and in the case of coarse signal sampling interval.

作者简介

E. Bazulin

ECHO+ Scientific Production Association

编辑信件的主要联系方式.
Email: bazulin@echoplus.ru
俄罗斯联邦, Moscow, 123458


版权所有 © Pleiades Publishing, Ltd., 2017
##common.cookie##