Modified carbon paste electrode for potentiometric determination of aluminium ion in spiked real water sample


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

The present paper describes the development of a new chemically modified carbon paste electrode based on azithromycin (AMC) that considered as a selective aluminium recognition agent in the carbon paste electrode (CPE). This electrode was fully characterized in terms of composition, response time, usable pH range and temperature. The electrode exhibited a Nernstian response for Al3+ ion over a concentration range from 7.0 × 10–6 to 1.0 × 10–2 mol L–1 with a slope of 21.3 ± 0.18 mV decade–1 and the limit of detection was 6.0 × 10–6 mol L−1. It had a response time of about 6 s. The proposed sensor possesses a very good selectivity with respect to a variety of other cations. Finally this modified electrode was applied for the determination of the concentration of Al3+ ion in different water samples and the obtained results were comparable with those obtained with atomic absorption spectrometer (AAS).

About the authors

Marwa ElBadry Mohamed

Chemistry Department, Faculty of Science

Author for correspondence.
Email: marelbadry@gmail.com
Egypt, Cairo, 12613


Copyright (c) 2016 Pleiades Publishing, Ltd.

This website uses cookies