Development of Digital MRI Consoles Using General-Purpose Digital Instruments and Microcontroller Boards


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

We developed two digital magnetic resonance imaging (MRI) consoles using general-purpose digital instruments and microcontroller boards. The first console consisted of a digital oscilloscope (8-bit resolution, 250-MHz sampling frequency), an arbitrary waveform generator (14-bit resolution, 100-MHz sampling frequency), and three 32-bit microcontroller boards. The second console consisted of a digital oscilloscope (16-bit resolution, 1-GHz sampling frequency) with a built-in waveform generator (14-bit resolution, 200-MHz sampling frequency) and three 32-bit microcontroller boards. MRI experiments were performed using a 1.0-T and 90-mm gap yokeless permanent magnet to evaluate the MRI consoles. Three-dimensional spin-echo and gradient-echo images were successfully acquired using the first and second MRI consoles using an undersampling technique and RF phase correction. We concluded that the digital MRI consoles could be built using general-purpose digital instruments and microcontroller boards at a reduced cost and within a short development time.

作者简介

Katsumi Kose

Institute of Applied Physics, University of Tsukuba

编辑信件的主要联系方式.
Email: kose@bk.tsukuba.ac.jp
日本, Tsukuba, 305-8573

Makoto Tsuda

Institute of Applied Physics, University of Tsukuba

Email: kose@bk.tsukuba.ac.jp
日本, Tsukuba, 305-8573

Daiki Tamada

Institute of Applied Physics, University of Tsukuba

Email: kose@bk.tsukuba.ac.jp
日本, Tsukuba, 305-8573

Yasuhiko Terada

Institute of Applied Physics, University of Tsukuba

Email: kose@bk.tsukuba.ac.jp
日本, Tsukuba, 305-8573

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer-Verlag Wien, 2016