Transfer of the Temperature of the Triple Point of Water with an Acoustic Resonator Using the Comparison Method


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

We consider a method of transferring the temperature of the triple point of water with an acoustic resonator configuration for measuring the Boltzmann constant. It is shown that the transfer of the temperature of the triple point of water in a resonator using a highly sensitive differentiator does not require the use of absolute means of measurement and implemented with uncertainty limited only by thermal noise. The method enables reducing uncertainty in the temperature transfer when measuring the Boltzmann constant, whose accurate measurement is necessary for transition to a new definition of the unit of temperature.

About the authors

V. M. Malyshev

All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)

Email: pilipenko@vniiftri.ru
Russian Federation, Mendeleevo, Moscow Region

K. D. Pilipenko

All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)

Author for correspondence.
Email: pilipenko@vniiftri.ru
Russian Federation, Mendeleevo, Moscow Region

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2017 Springer Science+Business Media New York