Use of the Four-Point Method for Measuring the Homogeneity of the Coating Deposited Inside Ceramic Vacuum Chambers of Pulsed Magnets for the MAX-IV Synchrotron Radiation Facility
- 作者: Anchugov O.V.1, Shvedov D.A.1
- 
							隶属关系: 
							- Budker Institute of Nuclear Physics, Siberian Branch, Russian Academy of Sciences
 
- 期: 卷 62, 编号 3 (2019)
- 页面: 440-443
- 栏目: Laboratory Techniques
- URL: https://journals.rcsi.science/0020-4412/article/view/160773
- DOI: https://doi.org/10.1134/S0020441219030023
- ID: 160773
如何引用文章
详细
An instrument has been developed and manufactured to measure the uniformity of the layer thickness deposited inside ceramic vacuum chambers of electron-beam injection devices (kickers). Using this instrument, it is possible to measure this parameter at an aperture size starting with 1 cm and at different aperture geometries. The measurement range is 0.1−100 µm. The necessity of such measurements is substantiated, various types of sensors are described, and results for some vacuum chambers of the injection kickers in the MAX-IV synchrotron radiation facility (Sweden) are presented.
作者简介
O. Anchugov
Budker Institute of Nuclear Physics, Siberian Branch, Russian Academy of Sciences
														Email: D.A.Shvedov@inp.nsk.su
				                					                																			                												                	俄罗斯联邦, 							Novosibirsk, 630090						
D. Shvedov
Budker Institute of Nuclear Physics, Siberian Branch, Russian Academy of Sciences
							编辑信件的主要联系方式.
							Email: D.A.Shvedov@inp.nsk.su
				                					                																			                												                	俄罗斯联邦, 							Novosibirsk, 630090						
补充文件
 
				
			 
						 
						 
					 
						 
						 
				 
  
  
  
  
  电邮这篇文章
			电邮这篇文章  开放存取
		                                开放存取 ##reader.subscriptionAccessGranted##
						##reader.subscriptionAccessGranted## 订阅存取
		                                		                                        订阅存取
		                                					