A dual absolute pressure measuring transducer
- Autores: Shtyrkov O.V.1, Yushkov V.A.1
- 
							Afiliações: 
							- Central Aerological Observatory
 
- Edição: Volume 59, Nº 1 (2016)
- Páginas: 139-141
- Seção: Laboratory Techniques
- URL: https://journals.rcsi.science/0020-4412/article/view/158921
- DOI: https://doi.org/10.1134/S0020441216010140
- ID: 158921
Citar
Resumo
The design of an ionization-type pressure measuring transducer is described. The meter has two output channels: for electron and ion currents. The conversion range varies from 10–3 to 10 Torr. The calibration characteristics are in good agreement with the predicted data of the classical impact-ionization theory. The standard calibration error is 7%. The calibrations of the meter were confirmed at the “Rostest-Moscow” Federal Budget Institution. To check the possibility of using the transducer for studying the middle atmosphere using meteorological rockets, the bench tests of the transducer for shock loads of up to 200 g were performed.
Sobre autores
O. Shtyrkov
Central Aerological Observatory
							Autor responsável pela correspondência
							Email: olegvalsh@yandex.ru
				                					                																			                												                	Rússia, 							ul. Pervomaiskaya 3, Dolgoprudnyi, Moscow oblast, 141700						
V. Yushkov
Central Aerological Observatory
														Email: olegvalsh@yandex.ru
				                					                																			                												                	Rússia, 							ul. Pervomaiskaya 3, Dolgoprudnyi, Moscow oblast, 141700						
Arquivos suplementares
 
				
			 
						 
						 
						 
						 
					 
				 
  
  
  
  
  Enviar artigo por via de e-mail
			Enviar artigo por via de e-mail  Acesso aberto
		                                Acesso aberto Acesso está concedido
						Acesso está concedido Somente assinantes
		                                		                                        Somente assinantes
		                                					