Measurements of Ion Beam Composition in the Temporary Radiobiology Facility at the Institute for High Energy Physics


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

A detector consisting of four p-i-n silicon pad structures of size 6 × 6 mm2 and thickness 300 μm was developed for the rapid monitoring of carbon ion beam quality at the Temporary Radiobiology Facility at the Institute for High Energy Physics (TRBF IHEP) by measuring event-by-event energy release on passage of charged particles (ions) through it. This articles describes the detector design and the method used for identifying contaminants in the beam of carbon nuclei. The charge composition of the ion beam of the TRBF IHEP calculated using the amplitude spectra of signals from the p-i-n structures is presented.

作者简介

A. Vorob’ev

A. A. Logunov Institute for High Energy Physics, National Research Center “Kurchatov Institute”

Email: pmk@ihep.ru
俄罗斯联邦, Protvino

A. Voronin

D. V. Skobeltsyn Institute of Nuclear Physics, M. V. Lomonosov Moscow State University

Email: pmk@ihep.ru
俄罗斯联邦, Moscow

S. Gorokhov

A. A. Logunov Institute for High Energy Physics, National Research Center “Kurchatov Institute”

Email: pmk@ihep.ru
俄罗斯联邦, Protvino

A. Kiryakov

A. A. Logunov Institute for High Energy Physics, National Research Center “Kurchatov Institute”

Email: pmk@ihep.ru
俄罗斯联邦, Protvino

M. Merkin

D. V. Skobeltsyn Institute of Nuclear Physics, M. V. Lomonosov Moscow State University

Email: pmk@ihep.ru
俄罗斯联邦, Moscow

M. Polkovnikov

A. A. Logunov Institute for High Energy Physics, National Research Center “Kurchatov Institute”

编辑信件的主要联系方式.
Email: pmk@ihep.ru
俄罗斯联邦, Protvino

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer Science+Business Media, LLC, part of Springer Nature, 2019