Analysis of the Preload and Afterload Sensitivity of the Sputnik Rotary Blood Pump


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The reaction of the Sputnik rotary blood pump to preload and afterload was analyzed. A mathematical model of the blood pump was suggested. The characteristics of the Sputnik pump were compared to those of the DuraHeart and INCOR commercial pump models. The effect of afterload on the pump capacity and preload sensitivity was analyzed. The preload sensitivity was shown to correlate with the shape of the head–capacity curves. It was demonstrated that the Sputnik pump had higher sensitivity to preload [mean value, (0.121 ± 0.0092) l·min–1·mm Hg –1] because of the flatter head–capacity curve as compared to the DuraHeart and INCOR commercial pumps.

作者简介

D. Petukhov

National Research University of Electronic Technology

Email: telyshev@bms.zone
俄罗斯联邦, Zelenograd, Moscow

D. Telyshev

National Research University of Electronic Technology

编辑信件的主要联系方式.
Email: telyshev@bms.zone
俄罗斯联邦, Zelenograd, Moscow

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer Science+Business Media New York, 2016